Please use this identifier to cite or link to this item: https://hdl.handle.net/2440/23230
Citations
Scopus Web of Science® Altmetric
?
?
Type: Journal article
Title: Aging performance of a compliant layer bearing acetabular prosthesis in an ovine hip arthroplasty model
Author: Carbone, A.
Howie, D.
Mc Gee, M.
Field, J.
Pearcy, M.
Smith, N.
Jones, E.
Citation: Journal of Arthroplasty, 2006; 21(6):899-906
Publisher: Churchill Livingstone Inc Medical Publishers
Issue Date: 2006
ISSN: 0883-5403
1532-8406
Statement of
Responsibility: 
Angelo Carbone, Donald W. Howie, Margaret McGee, John Field, Mark Pearcy, Nigel Smith and Eric Jones
Abstract: The wear performance of a polyurethane compliant layer (CL) material formed into an acetabular component and implanted into a sheep model of cemented total hip arthroplasty was assessed at 6, 12, 24, and 48 months. Four (11%) of 36 acetabular components debonded from the cement and one component was slightly loose at the cement-bone interface. There was no macroscopic evidence of fracture, wear, or deformation of the CL material on the articular surface of the acetabular components. Small numbers of polymeric wear particles was found in the hip synovial tissues of 10 sheep, most commonly in the early time groups, and were likely associated with initial wear of surface asperities. The wear performance of the CL was unchanged during a 48-month implantation period.
Keywords: hip
arthroplasty
polyurethane
wear
in vivo
Description: Copyright © 2007 Elsevier B.V.
DOI: 10.1016/j.arth.2005.07.023
Description (link): http://www.elsevier.com/wps/find/journaldescription.cws_home/623141/description#description
Published version: http://dx.doi.org/10.1016/j.arth.2005.07.023
Appears in Collections:Aurora harvest 2
Orthopaedics and Trauma publications

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.