Please use this identifier to cite or link to this item: https://hdl.handle.net/2440/73387
Citations
Scopus Web of Science® Altmetric
?
?
Full metadata record
DC FieldValueLanguage
dc.contributor.authorHowie, D.-
dc.contributor.authorHolubowycz, O.-
dc.contributor.authorMiddleton, R.-
dc.date.issued2012-
dc.identifier.citationJournal of Bone and Joint Surgery: American Volume, 2012; 94(12):1095-1102-
dc.identifier.issn0021-9355-
dc.identifier.issn1535-1386-
dc.identifier.urihttp://hdl.handle.net/2440/73387-
dc.description.abstractThe use of larger femoral heads has been proposed to reduce the risk of dislocation after total hip arthroplasty, but there is a lack of evidence to support this proposal. The aim of this multicenter randomized controlled trial was to determine whether the incidence of dislocation one year after total hip arthroplasty is significantly lower in association with the use of a 36-mm femoral head articulation as compared with a 28-mm articulation.Six hundred and forty-four middle-aged and elderly patients undergoing primary or revision arthroplasty were randomized intraoperatively to receive either a 36 or 28-mm metal femoral head on highly cross-linked polyethylene. Patients who were at high risk of dislocation (including those with dementia and neuromuscular disease) and those undergoing revision for the treatment of recurrent hip dislocation or infection were excluded. Patients were stratified according to other potential risk factors for dislocation, including diagnosis and age. Diagnosis of hip dislocation required confirmation by a physician and radiographic evidence of a dislocation.Overall, at one year of follow-up, hips with a 36-mm femoral head articulation had a significantly lower incidence of dislocation than did those with a 28-mm articulation (1.3% [four of 299] compared with 5.4% [seventeen of 316]; difference, 4.1% [95% confidence interval, 1.2% to 7.2%]) when controlling for the type of procedure (primary or revision) (p = 0.012). The incidence of dislocation following primary arthroplasty was also significantly lower for hips with a 36-mm femoral head articulation than for those with a 28-mm articulation (0.8% [two of 258] compared with 4.4% [twelve of 275]; difference, 3.6% [95% confidence interval, 0.9% to 6.8%]) (p = 0.024). The incidence of dislocation following revision arthroplasty was 4.9% (two of forty-one) for hips with a 36-mm articulation and 12.2% (five of forty-one) for hips with a 28-mm articulation; this difference was not significant with the relatively small sample size of the revision group (difference, 7.3% [95% confidence interval, -5.9% to 21.1%]) (p = 0.273).Compared with a 28-mm femoral head articulation, a larger 36-mm articulation resulted in a significantly decreased incidence of dislocation in the first year following primary total hip arthroplasty. However, before a 36-mm metal-on-highly cross-linked polyethylene articulation is widely recommended, the incidence of late dislocation, wear, periprosthetic osteolysis, and liner fracture should be established.-
dc.description.statementofresponsibilityDonald W. Howie, MBBS, FRACS, PhD, Oksana T. Holubowycz, PhD, MPH, Robert Middleton, MBBChir, MA, FRCS(Orth), and the Large Articulation Study Group-
dc.language.isoen-
dc.publisherJournal Bone Joint Surgery Inc-
dc.rightsCopyright © 2012 by The Journal Of Bone And Joint Surgery, Incorporated-
dc.source.urihttp://jbjs.org/article.aspx?articleid=1181958-
dc.subjectLarge Articulation Study Group-
dc.subjectFemur Head-
dc.subjectHumans-
dc.subjectOsteoarthritis, Hip-
dc.subjectHip Dislocation-
dc.subjectProsthesis Failure-
dc.subjectArthroplasty, Replacement, Hip-
dc.subjectIncidence-
dc.subjectRisk Factors-
dc.subjectCohort Studies-
dc.subjectProsthesis Design-
dc.subjectHip Prosthesis-
dc.subjectAged-
dc.subjectAged, 80 and over-
dc.subjectMiddle Aged-
dc.subjectFemale-
dc.subjectMale-
dc.titleLarge femoral heads decrease the incidence of dislocation after total hip arthroplasty: A randomized controlled trial-
dc.typeJournal article-
dc.identifier.doi10.2106/JBJS.K.00570-
pubs.publication-statusPublished-
dc.identifier.orcidHowie, D. [0000-0003-1702-3279]-
Appears in Collections:Aurora harvest 5
Orthopaedics and Trauma publications

Files in This Item:
File Description SizeFormat 
hdl_73387.pdfPublished version818.64 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.