dc.description.abstractWe have compared prospectively the incidence of loosening of 20 femoral stems with a matt surface with that of 20 polished stems of an otherwise identical tapered, non-modular design of Exeter hip replacement. The stems were inserted using the same technique at operation and radiographs showed no difference in the adequacy of the cement mantle or of fixation. All the patients were reviewed regularly and none was lost to follow-up. After a minimum follow-up of nine years, four matt but no polished stems had been revised for aseptic loosening. Polished stems subsided slightly within the cement mantle early, but did not loosen.-
dc.description.statementofresponsibilityDonald W. Howie, Robert G. Middleton, Kerry Costi-
dc.language.isoen-
dc.publisherBritish Editorial Society of Bone and Joint Surgery-
dc.rights© 1998 British Editorial Society of Bone and Joint Surgery-
dc.source.urihttp://www.bjj.boneandjoint.org.uk/content/80-B/4/573-
dc.subjectFemur-
dc.subjectAcetabulum-
dc.subjectHumans-
dc.subjectArthritis, Rheumatoid-
dc.subjectOsteoarthritis-
dc.subjectFemoral Neck Fractures-
dc.subjectProsthesis Failure-
dc.subjectAlloys-
dc.subjectRadiography-
dc.subjectArthroplasty, Replacement, Hip-
dc.subjectReoperation-
dc.subjectCementation-
dc.subjectIncidence-
dc.subjectFollow-Up Studies-
dc.subjectProspective Studies-
dc.subjectProsthesis Design-
dc.subjectHip Prosthesis-
dc.subjectSurface Properties-
dc.subjectAged-
dc.subjectAged, 80 and over-
dc.subjectMiddle Aged-
dc.subjectFemale-
dc.subjectMale-
dc.titleLoosening of matt and polished cemented femoral stems-
dc.typeJournal article-
dc.identifier.doi10.1302/0301-620X.80B4.8629-
pubs.publication-statusPublished-
dc.identifier.orcidHowie, D. [0000-0003-1702-3279]-
Appears in Collections:Aurora harvest 5
Orthopaedics and Trauma publications

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.