Please use this identifier to cite or link to this item: https://hdl.handle.net/2440/6682
Citations
Scopus Web of Science® Altmetric
?
?
Full metadata record
DC FieldValueLanguage
dc.contributor.authorMiddleton, R.-
dc.contributor.authorHowie, D.-
dc.contributor.authorCosti, K.-
dc.contributor.authorSharpe, P.-
dc.date.issued1998-
dc.identifier.citationClinical Orthopaedics and Related Research, 1998; 13(355):47-56-
dc.identifier.issn0009-921X-
dc.identifier.issn1528-1132-
dc.identifier.urihttp://hdl.handle.net/2440/6682-
dc.description.abstractThe effects of matte finish and modularity on loosening of tapered stems using the same cementing technique were studied prospectively. In 80 patients, 82 cemented Exeter primary stems were implanted at total hip revision by one surgeon using the same surgical and cementing technique throughout the series. The polished stems behaved differently than the matte surfaced stems behaved. Polished stems subsided in the cement mantle an average of 1 mm at 2 years after implantation, but without subsequent loosening of stems at as long as 12 years after implantation. Matte surfaced stems with metal centralizers had a higher loosening rate, and loss of fixation at the prosthesis to cement interface was identified as an early sign of loosening of these stems. At a mean 6-year followup, there were no revisions nor was there radiographic evidence of loosening of the polished modular stems. It is concluded that matte finish results in increased loosening of tapered stems but the introduction of modularity did not.-
dc.description.statementofresponsibilityMiddleton, Robert G.; Howie, Donald W.; Costi, Kerry; Sharpe, Phillipa-
dc.language.isoen-
dc.publisherLIPPINCOTT WILLIAMS & WILKINS-
dc.rights© 1998 Lippincott Williams & Wilkins, Inc.-
dc.source.urihttp://dx.doi.org/10.1097/00003086-199810000-00006-
dc.subjectHumans-
dc.subjectProsthesis Failure-
dc.subjectBone Cements-
dc.subjectArthroplasty, Replacement, Hip-
dc.subjectProspective Studies-
dc.subjectProsthesis Design-
dc.subjectEquipment Failure Analysis-
dc.subjectHip Prosthesis-
dc.subjectSurface Properties-
dc.subjectTime Factors-
dc.subjectAged-
dc.subjectAged, 80 and over-
dc.subjectMiddle Aged-
dc.titleEffects of design changes on cemented tapered femoral stem fixation-
dc.typeJournal article-
dc.identifier.doi10.1097/00003086-199810000-00006-
pubs.publication-statusPublished-
dc.identifier.orcidHowie, D. [0000-0003-1702-3279]-
Appears in Collections:Aurora harvest
Orthopaedics and Trauma publications

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.