dc.description.abstractWe present a 51-year-old morbidly obese man who underwent insertion of a single-chamber implantable cardioverter defibrillator for monomorphic ventricular tachycardia occurring after myocardial infarction. After a period of satisfactory device function, a sudden change in R-wave with atrial over-sensing heralded inappropriate defibrillation and induction of ventricular fibrillation, with subsequent death of the patient.-
dc.description.statementofresponsibilityHany Dimitri, Bobby John, Glenn D. Young and Prashanthan Sanders-
dc.language.isoen-
dc.publisherW B Saunders Co Ltd-
dc.source.urihttp://dx.doi.org/10.1093/europace/eum186-
dc.subjectVentricular tachycardia-
dc.subjectventricular fibrillation-
dc.subjectinappropriate shock-
dc.subjectfefibrillation-
dc.subjectimplantable cardioverter defibrillator-
dc.subjectlead dislodgement-
dc.titleFatal outcome from inappropriate defibrillation-
dc.typeJournal article-
dc.identifier.doi10.1093/europace/eum186-
pubs.publication-statusPublished-
dc.identifier.orcidSanders, P. [0000-0003-3803-8429]-
Appears in Collections:Aurora harvest
Medicine publications

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.