Please use this identifier to cite or link to this item:
https://hdl.handle.net/2440/42241
Citations | ||
Scopus | Web of Science® | Altmetric |
---|---|---|
?
|
?
|
Type: | Journal article |
Title: | Fatal outcome from inappropriate defibrillation |
Author: | Dimitri, H. John, B. Young, G. Sanders, P. |
Citation: | Europace, 2007; 9(11):1059-1060 |
Publisher: | W B Saunders Co Ltd |
Issue Date: | 2007 |
ISSN: | 1099-5129 1532-2092 |
Statement of Responsibility: | Hany Dimitri, Bobby John, Glenn D. Young and Prashanthan Sanders |
Abstract: | We present a 51-year-old morbidly obese man who underwent insertion of a single-chamber implantable cardioverter defibrillator for monomorphic ventricular tachycardia occurring after myocardial infarction. After a period of satisfactory device function, a sudden change in R-wave with atrial over-sensing heralded inappropriate defibrillation and induction of ventricular fibrillation, with subsequent death of the patient. |
Keywords: | Ventricular tachycardia ventricular fibrillation inappropriate shock fefibrillation implantable cardioverter defibrillator lead dislodgement |
Description: | © The European Society of Cardiology 2007. All rights reserved. |
DOI: | 10.1093/europace/eum186 |
Published version: | http://dx.doi.org/10.1093/europace/eum186 |
Appears in Collections: | Aurora harvest Medicine publications |