Title: Adapting detection sensitivity based on evidence of irregular sinus arrhythmia to improve atrial fibrillation detection in insertable cardiac monitors
Author: Pürerfellner, H.
Sanders, P.
Sarkar, S.
Reisfeld, E.
Reiland, J.
Koehler, J.
Pokushalov, E.
Urban, L.
Dekker, L.
Citation: Europace, 2017; 20(3):1-8
Publisher: Oxford University Press
Issue Date: 2017
ISSN: 1099-5129
1532-2092
Statement of
Responsibility: 
Helmut Pürerfellner, Prashanthan Sanders, Shantanu Sarkar, Erin Reisfeld, Jerry Reiland, Jodi Koehler, Evgeny Pokushalov, Ľuboš Urban, Lukas R C Dekker
Abstract: Intermittent change in p-wave discernibility during periods of ectopy and sinus arrhythmia is a cause of inappropriate atrial fibrillation (AF) detection in insertable cardiac monitors (ICM). To address this, we developed and validated an enhanced AF detection algorithm.Atrial fibrillation detection in Reveal LINQ ICM uses patterns of incoherence in RR intervals and absence of P-wave evidence over a 2-min period. The enhanced algorithm includes P-wave evidence during RR irregularity as evidence of sinus arrhythmia or ectopy to adaptively optimize sensitivity for AF detection. The algorithm was developed and validated using Holter data from the XPECT and LINQ Usability studies which collected surface electrocardiogram (ECG) and continuous ICM ECG over a 24-48 h period. The algorithm detections were compared with Holter annotations, performed by multiple reviewers, to compute episode and duration detection performance. The validation dataset comprised of 3187 h of valid Holter and LINQ recordings from 138 patients, with true AF in 37 patients yielding 108 true AF episodes ≥2-min and 449 h of AF. The enhanced algorithm reduced inappropriately detected episodes by 49% and duration by 66% with <1% loss in true episodes or duration. The algorithm correctly identified 98.9% of total AF duration and 99.8% of total sinus or non-AF rhythm duration. The algorithm detected 97.2% (99.7% per-patient average) of all AF episodes ≥2-min, and 84.9% (95.3% per-patient average) of detected episodes involved AF.An enhancement that adapts sensitivity for AF detection reduced inappropriately detected episodes and duration with minimal reduction in sensitivity.
Keywords: Atrial fibrillation
Diagnosis and monitoring
Insertable cardiac monitors
Rights: © The Author 2017. Published by Oxford University Press on behalf of the European Society of Cardiology. This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited. For commercial re-use, please contact journals.permissions@oup.com
DOI: 10.1093/europace/eux272
Published version: http://dx.doi.org/10.1093/europace/eux272
Appears in Collections:Aurora harvest 8
Medicine publications

Files in This Item:
File Description SizeFormat 
hdl_116469.pdfPublished version503.55 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.